ATR MWIR methodology
The laboratory measures an infrared absorption spectrum of the submitted sample and compares it with reference information.
Method overview and instrument configuration
| Instrument | ATR MWIR spectrometer |
|---|---|
| Range | 5.5–11 µm (approx. 1818–909 cm⁻¹) |
ATR principle
Attenuated total reflectance is an infrared sampling technique in which the sample is brought into contact with an internal reflection element. The beam is totally reflected inside the element, and the sample is probed through the resulting evanescent field. Absorption by the sample attenuates the reflected light at characteristic wavelengths.
Spectral acquisition details are described at a high level; SOP specifics are not published.
Reference comparison
The measured spectrum is compared with appropriate reference information to assess whether characteristic absorption features and the overall profile are congruent with the expected material. Reference creation and acceptance criteria are still being defined.
Interpretation framework
The measured spectral characteristics are consistent with the expected reference within the validated interpretation criteria.
The sample exhibits spectral differences from the expected reference. Additional analytical testing may be appropriate.
The measured spectrum is not congruent with the expected reference. Material spectral differences were observed.
Pass and Fail describe spectral congruence with the expected reference only. They are not a safety, potency or quality judgment. Criteria are subject to method validation.
Analytical limitations
ATR MWIR screening does not necessarily establish molecular identity, chromatographic purity, concentration, sterility, endotoxin content, biological activity, potency or safety, and may not detect every contaminant, adulterant, excipient or degradation product. Interpretation is limited to the spectral information within the measurement range.
Relationship to HPLC / MS
Validation status
Method validation is not yet complete. No sensitivity, specificity, accuracy or confidence figures are published. A validation summary will be added when study results are available.